4.4

Information on the measuring point

The following information on the measuring point can be called up with the handheld
terminal:
Parameter field
Display or entry
Measured values
V0H0
Main measured value for pressure
V7H0
Current display: Actual current in mA
V7H8
Sensor pressure (units in V0H9) selectable
V9H7
Actual dampened pressure without bias correction
Sensor data
V7H4
Lower calibration pressure
V7H5
Upper calibration pressure
V7H6
Lower measurement limit of sensor (units in V0H9 selectable)
V7H7
Upper measurement limit of sensor (units in V0H9 selectable)
Information on transmitter
V2H2
8010 = Software number
Sensor data No.: Number of entry in the sensor table (1...10).
V2H7
Please remove from sensor pass
Sensor data value: Entry in sensor table, contains all sensor-specific data. Please
V2H8
remove from sensor pass
Error response
V2H0
Actual diagnostic code
V2H1
Last diagnostic code
Parameter field
Display or entry
Measuring point tag. The measuring point can be identified here with a max. of 8
VAH0
characters.
VAH1
Descriptor. A max. of 16 characters can be entered here for the Descriptor.
VAH2
User text. A max. of 8 characters can be entered here.
VAH3
Serial number of the transmitter
VAH4
Serial number of the sensor
Barcon LHC/PPC (HART)
Commissioning
Communication level
29