Siemens SIREMOBIL Compact L Service Instructions Manual page 70

Table of Contents

Advertisement

6 - 18
Display of realtime clock
Display of rotated image
Monitor zoom
Monitor split
Monitor text
Maximum image number
Disable dose rate selection
in user setup
Unit serial number
Product name
Default Program changeable
Create External disk
Default LUT
Default edge
Software backward compatible Configure No.
Configure the following parameters in User setup for MEMOSKOP:
Operating mode Parameters
VCR output
Hardcopy Release
Text monitor
Split screen
Display of time
Set Date Time
SIREMOBIL Compact
System Manual
Replacing boards / replacing components
Configure Yes. Select No only upon customer request.
Configure Yes.
Configure No.
Configure Yes.
Configure Yes.
Configure 100.
Set to "No" if equipped with an area dose product mea-
surement device. Menu "Storage of Dose on disk" and
menu "Dose on Hardcopy output" in User setup are visi-
ble. Otherwise, set to "Yes".
Program the serial number of the SIREMOBIL Compact.
Do not enter the first zero.
SIREMOBIL Compact (default). If wrong, configure
"SIREMOBIL Compact".
Default: Yes.
Depending on country specific regulations configure Yes
or No.
Configure No.
Configure 2. Program a different LUT curve only upon
customer request.
Configure 1. Program a different Edge enhancement
value only upon customer request.
Change the programmed parameters only upon cus-
tomer request.
Configure to monitor A.
For SIREMOBIL Compact with 1 monitor:
configure to monitor A.
For SIREMOBIL Compact with 2 monitors:
Configure to monitor B.
For SIREMOBIL Compact with 1 monitor:
configure to monitor A.
For SIREMOBIL Compact with 2 monitors:
configure to both monitors.
Configure horizontally. Configure vertically only upon
customer request.
Configure Yes.
Set the current date and time.
SPR2-130.061.02
Rev. 11
Page 18 of 28
09.04
CS PS 24
Siemens AG
Medical Solutions