Measurement

Switch on fluoroscopy.
Read and record the value displayed on the dose rate meter.
If the value measured exceeds 9 R/min, the maximum tube current must be reduced.
If the value measured is less than 9 R/min, the tube current does not need to be reduced.
Switch fluoroscopy off.
Calculate the maximum tube current.
tube current displayed at 110kV in mA * 9 R/min
measured dose rate value in R/min
Multiply the tube current to be programmed by 10 , since entry on the service PC must be
made in 1/10mA. Round off the value to the nearest whole number.
Example: Tube current displayed at 110kV: 5mA
5 (mA) * 9 (R/min)
10.2 (R/min)

Programming

Connect the service PC to the serial interface of board D1.
Select the "Adjustment, Parameters" menu.
In the Combo box "Parameter Groups" select "Limitation of skin dose rate (FL)" .
Click on "Get from unit" .
Enter the value to be programmed in "Value" .
Click on "Set Value".
Select "Put to unit" . Read the menu guide on the service PC.
Select "Save to file".
The value to be programmed has been transferred to the SIREMOBIL Compact and
temporarily stored to the disk.
NOTE

Checking

After programming, check the maximum skin dose rate again.
SIREMOBIL Compact
System Manual
Maximum skin dose rate:
Measured dose rate value :
= 4.41 (mA)
You must exit the service program with "Logoff" and "Quit".
Only then will the modified parameters be stored on the floppy
disk as a new file.
Adjustments / Programs
= tube current to be programmed in mA
9R/min
10.2 R/min.
Rounded to the nearest whole number: 44
The value to be programmed is 44
SPR2-130.061.02
Rev. 11
09.04
Page 10 of 30
CS PS 24
5
5
5
Siemens AG
Medical Solutions