Line voltage adjustment
WARNING
•
Open the back cover of the monitor trolley.
•
Measure the on-site line voltage and line frequency.
•
Adjust transformer T1 to the required line voltage and line frequency. Refer to the label
on the switch-on assembly.
•
Adjust transformer T2 to the required line voltage and line frequency. Refer to the label
on the switch-on assembly.
•
Check fuses F1 and F2 to see if they correspond with the values on the label of the
switch-on assembly and replace them, if necessary.
•
Mark the correct line voltage and line frequency on the line voltage label at the back of
the monitor trolley.
Checking the line voltage on the monitor trolley
•
Connector X10 for the basic system must be disconnected.
•
The monitor, memory device and other devices, if present, such as multiformat camera,
video printer or video recorder must be disconnected from the line voltage supply.
•
Check the secondary voltages for line voltage transformer T1 according to Table 1.
from test point
Transformer T1.31
Transformer T1.31
Line voltage cable
Line voltage cable
Option:
Line voltage cable
Option:
Line voltage cable
Option:
Line voltage cable
Video printer, N
Option:
Line voltage cable
Video recorder, N
Option:
Optical isolation line
voltage cable, N
Tab. 1
Siemens AG
Medical Solutions
Electrical voltage!
See Chapter 1, Safety Information.
Disconnect the SIREMOBIL Compact line voltage plug.
Memoskop, N
Monitor 1, N
Monitor 2, N
Multispot, N
Optical isolation line L
to test point
Transformer T1.32
Transformer T1.33
Line voltage cable
Memoskop, L
Line voltage cable
Monitor 1, L
Line voltage cable
Monitor 2, L
Line voltage cable
Multispot, L
Line voltage cable
Video printer, L
Line voltage cable
Video recorder, L
voltage cable
SPR2-130.061.02
Rev. 11
09.04
Voltage
Comment
190 V ~ to
cannot be adjusted
205 V ~
230 V ~ to
cannot be adjusted
246 V ~
230 V ~ to
cannot be adjusted
246 V ~
230 V ~ to
cannot be adjusted
246 V ~
230 V ~ to
cannot be adjusted
246 V ~
230 V ~ to
cannot be adjusted
246 V ~
230 V ~ to
cannot be adjusted
246 V ~
230 V ~ to
cannot be adjusted
246 V ~
230 V ~ to
cannot be adjusted
246 V ~
Page 1 of 2
SIREMOBIL Compact
CS PS 24
2 - 1
2
2