Siemens SIREMOBIL Compact L Service Instructions Manual page 50

Table of Contents

Advertisement

5 - 28
Function:
Create External disk (1)
Enable Bridge (1)
Default programs change-
able
Default LUT
Default LUT for Subtraction
Default Edge
Software backward compa-
tible
Unit serial number
Disable doserate selection
in user Setup
Subtraction K-factor,
phase A
Subtraction K-factor,
phase B
Subtraction Time of phase A
Roadmap K-factor, phase A
(1) in menu Diagnostics
SIREMOBIL Compact
System Manual
Adjustments / Programs
Comments:
MEMOSKOP CE and MEMOSKOP CE-100:
Program the value to "No".
All other types of MEMOSKOP:
If a MOD drive is implemented and/or the option DICOM
Bridge is installed, program the value to "Yes".
MEMOSKOP CE and MEMOSKOP CE-100:
Program the value to "No".
All other types of MEMOSKOP:
If a MOD drive is implemented and/or the option DICOM
Bridge is installed, program the value to "Yes".
Default: Yes
The organ programs can be protected against changes by
setting "No".
After switch-on, the programmed brightness/contrast
curve (LUT) is selected automatically.
Default: LUT 2
After switch-on, the programmed brightness/contrast
curve (LUT) is selected automatically for the subtraction
mode, when equipped with a MEMOSKOP C-SUB.
Default: LUT 1
After switch-on, the programmed edge enhancement is
selected automatically. Default: Edge 1
Not implemented "Must be set to No".
Enter the serial number of the SIREMOBIL Compact if it is
equipped with a Dicom bridge. The serial number is
required for identifying the images transferred. Please
note that you have to omit the leading 0 when entering the
serial number.
Set to "Yes" for SIREMOBIL Compact without an area
dose product measurement device.
Set to "No" for SIREMOBIL Compact with an area dose
product measurement device.
Noise suppression, phase A, is programmed in the SUB-
mode.
Phase A: Mask acquisition.
Noise suppression, phase B, is programmed in the SUB
mode.
Phase B: Acquisition of fill image.
The time of phase A in the SUB-mode is programmed.
Phase A: Mask acquisition.
Noise suppression, phase A, is programmed in the road-
map mode.
Phase A: Mask acquisition.
SPR2-130.061.02
Rev. 11
09.04
Page 28 of 30
CS PS 24
Siemens AG
Medical Solutions