Adjustments / Programs

Dose rate adjustment

WARNING

Prerequisites

The "Adjustment mode" parameter in the "Doserate adjustment" parameter group has
been programmed to "semi-automatic".

Preliminary steps

Switch on SIREMOBIL Compact.
Connect the service PC to the serial interface of the SIREMOBIL Compact.
Start the service program.
Select the "Diagnostics..." menu.
Select the "Monitoring..." menu.
Select the "Dose rate adjustment" menu from the Combo Box "Group 1".
Click on the button "UPDATE START".
The following is displayed on the service PC:
AUTO REGULATION
BRIGHTNESS DEVIATION shows the difference in the actual brightness value minus the
CONTROL CURVE
ADJUSTMENT
I.I. ZOOM STATUS
DOSERATE
TV-IRIS NOMINAL VALUE shows the stored or nominal TV-iris value from the learning
TV-IRIS ACTUAL VALUE
Insert 25 mm Al and 2.1 mm Cu at the radiation exit port of the SIREPHOS.
The grid remains on the I.I.
Siemens AG
Medical Solutions
X-ray radiation!
See Chapter 1, Safety Information.
Protect yourself against radiation exposure. Wear a lead apron.
indicates the current status of the automatic adjustment.
(active / not active / manual / done)
nominal brightness value.
shows the currently selected fluoro curve (default: S2, antiiso-
watt 5 mA)
shows the operating control mode.
(Generator / Camera Iris / Stop)
shows the current I.I. format (survey / zoom)
shows the dose rate selected.
(LOW / MID / HIGH)
phase.
Note: An offset value is added to this value.
For this reason, the TV-IRIS NOMINAL VALUE and TV-IRIS
ACTUAL VALUE differ despite the correct adjustment
(BRIGHTNESS DEVIATION 0 + 3).
shows the actual TV iris value.
SPR2-130.061.02
Rev. 11
Page 5 of 30
09.04
CS PS 24
5 - 5
5
5
5
SIREMOBIL Compact