Concluding Steps; Checking The Dose Rate In Automatic Mode - Siemens SIREMOBIL Compact L Service Instructions Manual

Table of Contents

Advertisement

5 - 8
Select the next level of dose rate (LOW / MID / HIGH) and repeat the dose rate setting for
the survey and zoom format for this level.

Concluding steps

After setting all dose rates for the survey and zoom formats, click "Stop Update" and wait
until "CAL" disappears from the 7 segment indicator to the right on the operating con-
sole. The TV- iris collimator values are stored to the EE-PROM of the host.
Press ESC at the service PC to return to the main menu.
Select the "Data, Backup..." menu.
Click "Parameter" in the backup window.
- Then click "Backup" .
Enter" dose rate set" under "Remarks for Parameter".
Then click "OK" . The parameters are stored to disk.
You can now exit from the service program.

Checking the dose rate in automatic mode

NOTE
Center and reattach the dose measurement chamber to the I.I. input screen.
Set the kV-mA curve HC1 (3 mA plateau curve). If necessary, temporarily program the
HC1 curve.
During automatic mode (kV STOP is off), check the dose rate for dose rate levels
LOW / MID / HIGH for the survey and zoom format.Take into account the correction fac-
tors for the grid and the dose measurement chamber. If necessary, temporarily change
the dose rate levels and kV-mA curves.
Tolerance margin for the dose rates: + 15%.
Set the kV-mA curve HC2 (5 mA plateau curve). If necessary, temporarily program the
HC2 curve.
During automatic mode (kV STOP is off), check the dose rate for dose rate levels
LOW / MID / HIGH for the survey and zoom format.Take into account the correction fac-
tors for the grid and the dose measurement chamber. If necessary, temporarily change
the dose rate levels and kV-mA curves.
Tolerance margin for the dose rates: + 15%.
SIREMOBIL Compact
System Manual
Checking the dose rate in zoom format with a dose rate measure-
ment chamber that is not completely irradiated may lead to incor-
rect measurements.
SPR2-130.061.02
Rev. 11
Adjustments / Programs
Page 8 of 30
09.04
CS PS 24
5
5
Siemens AG
Medical Solutions