It is convenient to detach and attach instrument by
loosening or tightening the locking lever.
Detachment:
a. Turn locking lever 180° in counter clock-wise
direction.
b. Lift the instrument up with one hand carrying
handle and another hand holding the tribrach.
Attachment:
a. Match the instrument base with the correct groove
before putting the instrument on the board.
b. Tighten the locking lever
instrument base
groove
locking lever

Error display

E01
Vertical angle 0 position is out of range or set with
incorrect procedure.
E02
Tilt angle compensator 0 position is out of range or
set with incorrect procedure.
E03
During measuring of the collimation error, the value
measured is out of range.
E04
There's abnormality in internal memory system.
E05
Reserved for adjustment in factory.
E06
There's abnormality in angle measuring system.
E07
The level collimation or the telescope revolves too
fast (over 4 r/s).
E08
There's a error detected in angle measuring system.
The instrument should be re-powered to eliminate
this error.