Significant Event Log: Event IDs (Continued)
Code Short text for code
308
SID POST MAX SYSTEM
RESETS
309
SID POST INDICATOR FAIL
310
SID POST ABNORMAL
RESTAR
332
SID POST VBATT MONITOR
FAIL
Table 5-3: Significant Event Log: Event IDs (Sheet 14 of 22)
BiPAP® Focus™ Ventilator Service Manual © Respironics, Inc.
Comment / Corrective Action
The system has detected either, 5 soft resets early in
the UI processor POST or the system declaring a
system error because 3 resets occurred in the last 24
hours.
1. Examine the significant event log, check the recent
error codes and see if the Entries point to either a
Controller PCB, Controller subassembly or UI PCB
issue.
2. If this persists replace the UI PCB
3. If replacing the UI PCB doesn't fix the issue, replace
the controller PCB.
UI processor detected one of the leds is faulty during
POST
1. Enter Diagnostics mode and use the System Test
screen, to test the leds and identify which led is
faulty.
2. If no led is found to be faulty and If this persists and
leads to multiple resets and the system declaring a
system error because too many errors occurred in 24
hours, replace the UI PCB.
The system has detected that a soft reset has occurred
unexpectedly.
1. Examine the significant event log, check the recent
error codes and see if the entries point to either a
Controller PCB, Controller subassembly or UI PCB
issue.
2. If this persists replace the UI PCB
3. If replacing the UI PCB doesn't fix the issue, replace
the controller PCB.
UI processor detected that the VBATT test failed during
POST.
1. If running on battery, ensure the system has a
charged battery as per the setup section in the
Operator's Manual.
2. Check if the problem exists when POST is run on
mains and battery.
3. Enter Diagnostics mode and use the Sensor
Readings screen, to check the adc readings.
4. If this persists and leads to multiple resets and the
system declaring a system error because too many
errors occurred in 24 hours, replace the UI PCB.