Chapter 5
Diagnostic Mode and Troubleshooting
5-14
Diagnostics Label
Actual Measured Value
PWRSWMON
PWR_SW_MON_PROC
PWRSWLATCH
PWR_SW_LATCHED
Table 5-1: Sensor Readings Defined (Sheet 4 of 4)
BiPAP® Focus™ Ventilator Service Manual © Respironics, Inc.
Sensor Readings Defined
Description
User interface Standby Key monitor
Hi: Standby key is in-active (released).
Lo: Standby key is active (pressed).
Latched User interface Standby Key
monitor. This signal must be primed by
software with the PWR_SW_CLK signal.
This can be done using the PWRSWCLK
diagnostics label in the System Test screen.
Hi: Standby key has NOT been pressed.
Lo: Standby key has been pressed.
REF 1029568 Rev A