Supplemental power socket

(reference 30400-01)

Available exclusively and standard EU & UK
Safety note
When a medical device is connected to the supplemental power socket, the system
becomes an medical electrical device in accordance with Standard EN 60601-1. It is
the responsibility of the user to ensure compliance of the medical electrical system
with standard EN 60601-1 (Article 16)

Using the supplemental power socket

• Properties :
The power consumed in a single event must not exceed 600VA.
Options 30400-01 (supplemental power socket) consists of 2 sockets, one on each side of the base of the
couch.
When simultaneously using products 30400-01 & 3058-01 (supplemental power
socket and the circuit breaker), the sockets remain continuously powered, even if the
safety key is withdrawn.
The electrical sockets are equipped with hinged covers.
When these covers are closed, the examination couch
is protected against sprays of water (IP X4 according to
European standard 60 529). If one of the covers is open
(i.e. each time one the sockets is in use), this protection is
no longer present.
User's Guide
120 V or 230 V depending on the country
2.6A AC max
x2