Error Limits - BRAND Dispensette TA Operating Manual

Trace analysis
Hide thumbs Also See for Dispensette TA:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 25

Error limits

Error limits according DIN EN ISO 8655-5 related to the nominal capacity
(= maximum volume) indicated on the instrument, obtained when instrument and
distilled water are equilibrated at ambient temperature (20 °C/68 °F). Testing
takes place according DIN EN ISO 8655-6 with a completely filled instrument
and with uniform and smooth dispensing.
Error limits Dispensette
Nominal volume
ml
10
* A = Accuracy, CV = Coefficient of Variation
Partial volume
The percentage values for A and CV are relative to the nominal volume (V
and must be converted for partial volumes (V
V
N
A
=
· A
V
T
N
T
e.g.
volume
V
N
V
= 50% N
T
V
= 10% N
T
* A = Accuracy, CV = Coefficient of Variation
Note:
The error limits are well within the limits of DIN EN ISO 8655-5. The maximum error limit for a single
measurement can be calculated EL = A + 2 CV (for volume 10 ml: 50 µl + 2 x 10 µl = 70 µl).
36
TA
®
A* ≤ ±
%
µl
0.5
50
A* ≤ ±
%
µl
10.0
0.5
50
5.0
1.0
50
1.0
5.0
50
CV* ≤
%
0.1
10
).
T
CV* ≤
%
0.1
0.2
1.0
µl
)
N
µl
10
10
10
20 °C
!
Ex

Advertisement

Table of Contents
loading

This manual is also suitable for:

4740 0404740 0414740 2404740 241

Table of Contents