3.6.2
Connecting Horizontal Arm Cables
1.
Connect the cable H001624 with the Connector 1, 2 on the Power Box Board, and the
Protective Earth (PE) Cable on the PE Conductor.
2.
Cover the Power Box Base with the Power Box Cover, and tighten two bolts with the
Power Box Deco.
EzRay Air W (Model: VEX-S300W) Installation Manual
1. Cable Name: H001624 (L=450 mm) / H001625 (L=600 mm)
H001626 (L=900 mm)
2. The installation procedures are the same, regardless of the
horizontal arm length.
CN (Connector) 2
3. Installing the Equipment
Protective Earth (PE) Cable
CN (Connector) 1
Power Box Deco