Spirometry setup

Setting patient and
sensor type
WARNING
Selecting a data source
3-26
2. Select Scaling.
3. Set the scale type to AUTO, Linked, or Indep.
Normal Screen
4. Push the
Use the Spirometry Setup menu to:
Set the patient and sensor type.
Select a data source.
Set the spirometry loop type.
Change the volume shown on the spirometry split screen to
MVexp or TVexp.
Set split-screen view.
Access the Paw and MVexp alarm limits.
Spirometry Setup
Patient and Sensor Type Adult
Data Source Vent
Loop Type P-V
TV or MV TV
Split Screen None
Paw Alarm
MVexp Alarm
Previous Menu
Patient and sensor type refer to the style of airway adapter used with
the airway module. If spirometry data is obtained from the airway
module, make sure that the sensor type matches the type airway
adapter used.
Make sure that the set sensor type corresponds to the
type of airway adapter in use. If the sensor type is not set
correctly, the information displayed may not be accurate.
Spirometry
1. Push the
2. Select Spiro Setup - Patient and Sensor Type.
3. Select Adult or Pedi depending on the sensor used.
Adult refers to the D-lite sensor.
Pedi refers to the Pedi-lite sensor.
Several monitoring parameters can be obtained from the ventilator or
from the airway module.
If information is not available through the airway module, information
comes from the internal ventilator sensors. If the internal ventilator
sensor readings disagree with the airway module sensor readings,
the associated alarm occurs.