Iqc; System Monitoring - Macherey-Nagel NANOCOLOR Advance Operating Instruction

Hide thumbs Also See for NANOCOLOR Advance:
Table of Contents

Advertisement

System

13.7 IQC

13.7.1 System Monitoring

13.7.1.1 System Check
13.7.1.2 Lamp Check
The Internal Quality Control (IQC) menu is accessed via the
icon. Through the various options, quality control parameter
can be checked and documented..
Clicking the icon
menu contains the programs for checking the optical properties of
the device. All tests can be carried out directly with the instrument
without additional material.
Clicking the icon
opens the dialog for the System Check menu.
The system check can be triggered manually by pressing
lamp test, wavelength accuracy test, filter test and detector test
are performed. This can take a few minutes. No measurements
are possible during this time. The result of the system check is
then displayed in the overview window.
If one or more tests fail, restart the device and repeat the mea-
surements. If the problems persist, contact the manufacturer or
your local distributor.
The Lamp Check serves to ensure that the installed halogen lamp
is working properly. To start the lamp check press
The device records a lamp curve and compares it with the stored
setpoints. The test is then evaluated. The result of the test can
then be saved to an external storage device in the form of a CSV
file by clicking on the
Spectrophotometer NANOCOLOR
opens the menu System Monitoring. The
icon.
Advance 1.0, 11.2020
®
. A
.
35