INSTALLATION INSTRUCTIONS

Filling the DEVELOPER TANK

[1] Check that the developer DRAIN VALVE is fully closed. Do not overtighten the VALVE.
Warning
• Wear rubber gloves, safety glasses, and protective clothing when mixing chemicals and filling the tanks.
• To prevent spills or splashes into other TANKS or onto electrical parts:
– Check that the SIDE ACCESS PANELS are installed.
– Pour the developer into the TANK at the non-drive side of the PROCESSOR.
– Do not pour developer onto the RACK.
[2] Mix the volume of developer solution required in the DEVELOPER TANK as recommended by the directions
packaged with the chemicals.
[3] Check that the DEVELOPER TANK does not contain any spills of fixer solution. If necessary, rinse and dry
the DEVELOPER TANK by wiping the TANK with a clean, lint-free cloth.
Checking the Level of the Developer Solution
Figure 61
HEAT
EXCHANGER
Important
It is important to do the following steps in the correct sequence to prevent Kodak RP X-Omat Developer Starter from
going down the DRAIN HOSES.
[4] Pour developer solution into the DEVELOPER TANK until the level of the solution completely covers the
HEAT EXCHANGER in the bottom of the DEVELOPER TANK. See Figure 61 on Page 70.
[5] Pour 266 mL (9 fl oz.) of Kodak RP X-Omat Developer Starter into the DEVELOPER TANK.
70
SPLASH GUARD
P/N 1C4019
September 1995 – 981088
STEP
DEVELOPER
TANK
developer
DRAIN VALVE
H130_0109HCE
H130_0109HA