Technical data
Dimensions (W x D x H)
Weight
Power supply
Mains power
DC power
Battery power (per battery)
Operating time under battery
power
1
point 2 CPAP and Auto-
CPAP
point 2 BILEVEL ST20/25
Running reserve of internal
clock
Pressure range
Max. pressure limit in the
event of a fault
Power consumption
Standby
(battery charging)
Standby (
without batteries)
Operation at 25 hPa
Operation at 20 hPa
Operation at 12 hPa
Operation at 6 hPa
Short term pressure variation point 2 CPAP, AutoCPAP, BILEVEL ST20 in hPa
10 bpm
15 bpm
20 bpm
The operating time is depending on the device settings and the age and stage of
1
charge of the batteries.
point 2 BILEVEL ST25
2
90
Annex
point 2
170 x 220 x 95 mm
1.5 kg
100 ... 240 V AC, 50 ... 60 Hz
24 V DC / max. 2.01 A
11.1 V / 2200 mAh
> 8 hours (at 10 hPa, 12 bpm,
500 ml tidal volume)
> 7 hours (at IPAP = 12 hPa,
EPAP = 6 hPa, f = 12 bpm,
300 ml tidal volume,
I:E = 40 % : 60 %, T mode)
4...20 (25
DC power
Mains power
< 26 W
< 35 W
< 3 W
< 5 W
< 17 W
< 21 W
< 14 W
< 17 W
< 10 W
< 12 W
< 6 W
< 9 W
4
8
12
16
hPa
hPa
hPa
hPa
0.15 0.15 0.17 0.20 0.25 0.16 0.19 0.30 0.40 0.48
0.23 0.24 0.27 0.32 0.38 0.24 0.31 0.45 0.58 0.72
0.36 0.32 0.37 0.43 0.49 0.32 0.42 0.62 0.70 0.91
point 2 with aquapoint 2
180 x 350 x 110 mm
approx. 1.9 kg
(without water)
up to 8 years
) hPa (mbar)
2
≤ 30 hPa
DC power
-
-
< 39 W
< 38 W
< 34 W
< 30 W
20
4
8
12
hPa
hPa
hPa
hPa
(at heating level 3)
-
-
Mains power
-
-
< 48 W
< 42 W
< 38 W
< 35 W
16
20
hPa
hPa