Electrical safety

• Do not try to open the device or the power supply unit. Maintenance and repairs
may only be performed by personnel authorized by HOFFRICHTER GmbH.
• Do not touch live parts of the power cable or power supply unit if they are faulty.
• The device must not be used in wet rooms, as there is a risk of electrical shock
from moisture in the device.
• Only the power supply unit supplied may be used for operating the device.
• Do not use any electrically conductive or electrostatically chargeable patient tubes.
• The device must not be closer than 30 cm to other devices such as defibrillators,
diathermy units, mobile phones, microwaves, remote controlled toys, etc. Elec-
tromagnetic fields that exceed 10 V/m may adversely affect the operation of the
ventilator.
• To disconnect the device from the power supply, the power plug must be pulled out.
• Do not put any containers which contain liquid on the device.
• Do not put the device near water containers (e.g. baths).
• The power plug must be unplugged before cleaning the device.
• Do not reach for the device under any circumstances should it fall into water.