Using Oxygen; Integration Into It Networks - Hoffrichter point 2 User Manual

Respiratory therapy device
Hide thumbs Also See for point 2:
Table of Contents

Advertisement

Using oxygen

• Please be sure to observe the user's manual of the manufacturer or distributor
from whom you obtain the oxygen.
• The patient must be advised about the use of oxygen.
• When oxygen is being used, appropriate training must be carried out in the home
environment.
• The distance between the oxygen source and the device must be at least 1 m.
• Oxygen supports combustion. Therefore, observe the fire protection regulations
applicable for using oxygen. Ensure that the oxygen fittings, as well as all ports
and surfaces near the oxygen lines, are free of grease. Do not smoke and do not
use naked flames. When using oxygen, an increased oxygen concentration in the
ambient air can occur.
• The oxygen supply should be stopped before the therapy is interrupted. We further
recommend that, after ending the therapy, the device is allowed to run for several
respiratory cycles without any oxygen supply.
• In the event of an oxygen leak, the oxygen source must be closed immediately.
The room must immediately be ventilated. At the same time, any sparks, flames
or potential sources of fire in the vicinity of the device must be avoided.

Integration into IT networks

• Integrating the device into an IT network that also includes other devices can lead
to previously unknown risks. The user is therefore responsible for assessing and
evaluating the risks and for risk minimization.
• Changes to the IT network can lead to new, unknown risks. This includes changes
to the network configuration, the integration and removal of elements as well as
performing updates and upgrades of devices in the IT network.
Safety Information
23

Hide quick links:

Advertisement

Table of Contents
loading