Microwave Output Connections (96270A)

To prevent hazardous RF transmissions and equipment
damage, follow the instructions in Chapter 3 before you make
connections to the microwave output or a UUT.
Instructions to connect the microwave output to a UUT are given in Chapter 3.
Do not make connections before you read all of the Cautions and Warnings
contained in these instructions.

Power Sensor Connections (96270A)

To prevent equipment damage, follow the instructions in
Chapter 3 before you connect the power sensor to the Product
or a UUT.
Instructions to connect a power sensor to the Instrument and to a UUT are given
in Chapter 3. Do not attempt to connect a power sensor before you read all of the
Cautions and Warnings contained in these instructions.

Install the Instrument in an Equipment Rack

The Instrument is suitable for both bench-top and rack-mounted operation. A
Rack Mount Slide Kit is available as an accessory. Instructions to install the kit
are supplied with the kit.

Cooling Considerations

Internal fans keep the operation temperature of the Instrument at a safe level. Air
intakes are on the left side and rear of the Instrument. Air is exhausted from the
right (as viewed from the front). For bench-top operation, these fans maintain
temperature control without any attention other than routine cleaning of the filter
as described in General Maintenance. However, when Instrument is mounted
and used in other situations (for example, in an equipment rack) additional
attention may be necessary to ensure that the Instrument is able to maintain a
normal operation temperature and does not overheat.
To prevent fire hazard and to ensure that the Product does not
exceed its normal operation temperature, observe these
warnings:
2-8
During normal operation, keep the Product covers securely
in place. Excessive air leaks can interrupt and redirect the
flow of cooling air from internal components.
When you mount the Product in an enclosed equipment
rack, provide adequate ventilation and airflow within the
rack. Pay particular attention to ensure adequate and proper
use of exhaust fans, louvers, equipment spacing, free-
flowing and isolated intake and exhaust ports.