Figure 3-52. Measure Profile Screen - Microwave Output Selected
Use Sensor 1 (Direct) for self-measurement of the Leveling Head or Microwave
Direct outputs with Sensor 1 connected as the reference sensor.
Use Sensor 2 (Direct) for self-measurement of the Leveling Head or Microwave
Direct outputs with Sensor 2 connected as the reference sensor.
Use Sensor 1 (Leveled) for self-measurement of the Microwave Sensor/Splitter
output with Sensor 1 connected as the reference sensor and Sensor 2 providing
leveling feedback.
Use Sensor 2 (Leveled) for self-measurement of the Microwave Sensor/Splitter
output with Sensor 2 connected as the reference sensor and Sensor 1 providing
leveling feedback.
To start the measurement process, push the Begin Measure softkey. When the
measurement process is started the Instrument initially uses the relevant power
sensor to check that the output of the Instrument is being applied to that power
sensor (via the cables/splitters/attenuators that are being characterized). Then
the Instrument goes through each point measuring and calculating the correction
factor. The points counter and progress bar is shown on the screen and are
updated accordingly.
The Instrument attempts to output sufficient signal to achieve
the demanded level at the reference sensor, which can be up to
the maximum Instrument output level. If the Instrument is
connected incorrectly, or if the attenuation to the measurement
plane (point of reference sensor connection) is larger than
anticipated connected, equipment may be damaged by
excessive levels.
Caution
3
Local Operation
Instrument Operation
hpn81.bmp
3-117