Cable Gland Assembly Instruction For Shielded Cables; Signal Inputs And Outputs - Emerson ROSEMOUNT MLT 2 Upgrade Manual

Pressurized analyzers intended to be used in hazardous areas
Hide thumbs Also See for ROSEMOUNT MLT 2:
Table of Contents

Advertisement

ATEX Instruction Manual
ETC01035
02/2008

2.7.1 Cable Gland Assembly Instruction for Shielded Cables

2.7.1
Cable Gland Assembly Instruction for Shielded Cables
1. Strip the cabel insu-
2. Uncover the shiel-
3.
4.
2.7.2

Signal Inputs and Outputs

Signal inputs and outputs by default are NOT
intrinsically safe!
Optional elements may be provided for in-
stallation on internal DIN rails, to convert
When the analyzer is out of order or if the pressurization unit shuts off
due to a failure all non-intrinsically safe inputs and outputs connected to
external equipment MUST be shut off too!
This ensures that no hazardous voltages are present within the analyzer
enclosure when not pressurized.
Emerson Process Management GmbH & Co. OHG
lation
ding
Feed cable through
gland nut and into
fixing element
Put the shielding net
o v e r t h e e l e m e n t
the way that it co-
vers the o-ring 2 mm
(0.08 in).
EXPLOSION HAZARD
MLT 2 / BINOS
5.
non-intrinsically safe signals to intrinsically
safe signals. Es können jedoch auf der im
Gerät vorhan
100 F
Stick the fixing ele-
ment into the neck
and fix the gland
2-17