Troubleshooting Prox Sensor - Veris 3150 MSP3 Operating Instructions Manual

Table of Contents

Advertisement

Procedure #14: Troubleshooting Prox Sensor
The prox sensor, located on the left of the sampler shoe parallel linkage arm, is critical to the
machine's function. It informs the external controller of the sampler's position, which allows the
controller to move the sampler to the next position in the sampling/measuring sequence. If the
prox sensor is not functioning properly, the entire operation will be affected. The problem may be
with the Prox sensor itself or with its adjustment. Follow flow chart below to address problems in
their logical sequence.
Figure 14.1.
1. Inspect LED light on prox sensor. Red LED light should light whenever prox sensor is
near metal and not light when away from metal (Figure 14.2). This sequence indicates to
the external controller the status of the shoe: whether the shoe is down and needs to come
up for pH measuring, or it's up and needs to go down for the next sample, etc. To view LED
light, shade ambient light from prox sensor and cycle sampler assembly manually. Be
careful to not strike or damage prox sensor face. KEEP HANDS CLEAR OF CYCLING
MECHANISM.
.
Figures 14.2a and b. Prox sensor and LED light
Prox Sensor
light
6-38
Pub. # OM20-MSP3-3D
Light is off when
sensor is above
and below plate,
and over center
gap; Light is on as
metal passes by
sensor

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

3210 msp3Msp3-3d

Table of Contents