15 Annex

15.1 Checklist before the operation

Patient
.......................................................................................................................................................................................
Surgeon
.......................................................................................................................................................................................
Date
.......................................................................................................................................................................................
Step Procedure
1
Cleaning the optical accessories
2
Installing the accessories
3
Checking the tube settings
4
Function check
5
Balancing
6
Sterility
7
Positioning at the OP table
72
Details
Check the tubes, eyepieces and the documentation
X
accessories (if used) for cleanliness.
Remove dust and dirt.
X
Lock the PROVIDO in place and install all accessories on the
X
microscope so it is ready for use (see page 24).
Position the handles as desired.
X
Connect the footswitch, if used.
X
Check the camera image on the monitor and realign if
X
necessary.
Check that all equipment is in its proper position (all
X
covers fitted, doors closed).
Check the tube and eyepiece setting for the selected user.
X
Check the fiber optics cable connection to the optics
X
carrier.
Connect the power cable.
X
Switch on the microscope.
X
Switch on the illuminator at the control unit.
X
Leave the illumination on for at least 5 minutes.
X
Check the lamp history and make sure that the remaining
X
life time is sufficient for the planned surgery.
Replace defective bulbs before the surgery.
X
Test all functions on the handles and the footswitch.
X
Check the user settings on the control unit for the selected
X
user.
Balance the PROVIDO (see page 13).
X
Press the "All Brakes" button on the handle and check the
X
balancing.
Fit sterile components and sterile drape if used (see
X
page 29).
Repeat balancing.
X
Position the PROVIDO on the OP table as required in
X
transport position and lock the foot brake after positioning
is done (see page 27).