16 Annex

16.1 Checklist before the operation

Patient
.......................................................................................................................................................................................
Surgeon
.......................................................................................................................................................................................
Date
.......................................................................................................................................................................................
Step Procedure
1
Cleaning the optical accessories
2
Installing the accessories
3
Checking the tube settings
4
Balancing
5
Function check
6
Safety check
7
Positioning at the OP table
8
Sterility
9
Final work
Details
Check the tubes, eyepieces and the documentation
X
accessories (if used) for cleanliness.
Remove dust and dirt.
X
Position the handles as desired.
X
Connect the mouth switch and/or footswitch if used.
X
Check the camera image on the monitor and realign if
X
necessary.
Check the tube and eyepiece setting for the selected user.
X
Balance the ARveo (see page 25).
X
Press the "All Brakes" button on the handle and check the
X
balancing.
Check the fiber optics cable connection to the optics
X
carrier.
Connect the power cable.
X
Switch on the microscope.
X
Switch on the illuminator at the control unit.
X
Leave the illumination on for at least 5 minutes.
X
Check the lamp history and make sure that the remaining
X
life time is sufficient for the planned surgery.
Replace defective bulbs before the surgery.
X
Test all functions on the handles and the footswitch.
X
Check the user settings on the control unit for the selected
X
user.
Check counterweights and accessories for secure seating.
X
Position the ARveo on the OP table as required and lock
X
the foot brake (see page 31).
Fit sterile components and sterile drape if used
X
(see page 32).
Check that all equipment is in its proper position
X
(all covers fitted, doors closed).
ARveo / Ref. 10 747 384 / Version 04
Annex
Checked / Signature
77