Positioning The Microscope; Adjusting The Microscope - Leica PROVIDO User Manual

Surgical microscope
Table of Contents

Advertisement

Operation
8.2

Positioning the microscope

8.2.1
Coarse positioning
Hold the microscope by both handles.
X
Press the button for releasing all brakes and position the
X
microscope.
Release the brakes button.
X
Also refer to the "Handles" chapter on page 22.
CAUTION
Damage to the PROVIDO due to uncontrolled tilting.
Hold the handle when releasing the brake.
X
8.2.2
Fine positioning (optional)
Position the microscope with the XY drive using the joystick on
X
the handle or the joystick on the footswitch.
You can change the speed at which the XY motors move on
the "Speed" menu screen
This value can be saved individually for each user
(see page 41).
32
.
PROVIDO Ref. 10732418 / Version 03
8.3

Adjusting the microscope

8.3.1
Adjusting the brightness
You can make the illumination brighter or darker using either the
touch panel monitor, a hand/footswitch, or handle.
On the touch panel monitor in the "Main" menu screen
Press the
or
button on the bar for adjusting the
X
brightness of the illumination "Lamp 1" or "Lamp 2".
– or –
Press the brightness adjustment bar directly.
X
The brightness of the active main illumination changes.
• Clicking the
value in increments of 1. Holding down the button with
your finger changes the value in increments of 5.
• The start setting can be saved individually for each user
(see page 41).
• The main illumination can only be switched on and off
using the illumination push-button on the stand.
• The brightness setting is also visible when the
illumination is off. However, the display bar will appear
darker.
On the footswitch/handle
Depending on the assignment (see page 44), you can also
increase and decrease the brightness of the main illuminator using
two correspondingly assigned buttons on the footswitch/handle.
or
button changes the brightness

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents