Attaching Sterile Controls And Drape - Leica PROVIDO User Manual

Surgical microscope
Table of Contents

Advertisement

564
625
PROVIDO Ref. 10732418 / Version 03
7.8
Attaching sterile controls and
drape
WARNING
Risk of infection.
Always use the PROVIDO with sterile controls and a
X
sterile drape.
7.8.1
Covers for rotary buttons
Use the covers also when you use sterile disposable drapes.
The controls will be easier to grasp.
Fit steam-sterilizable covers on the magnification, working
X
distance and AutoIris manual override knobs.
Attach steam-sterilizable covers to accessories as well (if
X
present).
7.8.2
Sterile drape for stand
• Only use the sterile Leica tested drapes specified in the
Accessories section.
• Drape only as far as the arm system (see picture below).
CAUTION
Risk of infection.
Leave sufficient space around the stand to ensure that
X
the sterile drape does not come into contact with non-
sterile components.
Activate the "All Brakes" function on the handle and extend the
X
arm system.
Put on sterile gloves.
X
Attach all the sterile controls.
X
Unpack the sterile drape carefully and drape it over the
X
Preparation before surgery