can significantly affect the measurement capabilities of the analyzer. To clean the
test ports (as well as other connectors of the analyzer), use the following
procedure:
 Using compressed air remove or loosen the contamination particles;
 Clean the connectors using a lint-free cleaning cloth wetted with a small
amount of ethanol and isopropyl alcohol (when cleaning a female
connector, avoid snagging the cloth on the center conductor contact
fingers by using short strokes);
 Dry the connector with low-pressure compressed air.
Always completely dry a connector before using it.
Never use water or abrasives for cleaning any connectors of the analyzer.
Do not allow contact of alcohol to the surface of the insulators of the
connectors.
WARNING
Never perform cleaning of the product if the power cable is
connected to the power outlet.
Never clean the internal components of the product.
 Storage
Before first use store your equipment in the factory package at environment
temperature from 0 to +50ºС and relative humidity up to 95% (at 25ºС).
After you have removed the factory package store the equipment at environment
temperature from +10 to +35ºС and relative humidity up to 80% (at 25ºС).
Ensure to keep the storage facilities free from dust, fumes of acids and alkalies,
aggressive gases, and other chemicals, which can cause corrosion.
© 2015 GS Instruments, Co., Ltd
SAFETY INSTRUCTIONS
7