Typical Setup For A Test Cell With Guard Ring - Omicron Lab SPECTANO 100 Quick Start Manual

Hide thumbs Also See for SPECTANO 100:
Table of Contents

Advertisement

To connect the adapter cable to SPECTANO 100:
1. Connect the standard triaxial cable according to the color code to the
2. Connect the adapter cables with the corresponding color code to the
7.4.1

Typical setup for a test cell with guard ring

Figure 7-4
SPECTANO 100 OUTPUT and INPUT connectors.
standard triaxial cable.
DANGER
Death or severe injury caused by high voltage or current
► Do not touch the OUTPUT connector when the red LED is on or blinks.
► Always ensure that no dangerous voltages are present at any touchable
part of the device, test cell or sample.
► Do not touch the output if the guard (inner shield of the triaxial connection)
is identical with the ground (outer shield of the triaxial connection).
Inner conductor
Outer (ground) &
inner (guard) shield
*
same potential
Connecting SPECTANO 100 to a test cell with guard ring
NOTICE
The safe area is only safe if the delivered cables and adapters are used.
Otherwise, this can also be a danger zone.
*
Outer (ground) &
inner (guard) shield
Inner conductor
Barrier
Installation
Electrodes
*
Guard ring
25