6

Device description

6.1

SPECTANO 100

6.1.1

Connectors and operating controls

Figure 6-1
DANGER
Death or severe injury caused by high voltage or current
► Do not touch the OUTPUT connector when the red LED is on or blinks.
► Always ensure that no dangerous voltages are present at any touchable
part of the device.
► Before handling SPECTANO 100 in any way, connect its equipotential
ground terminal with a solid connection of at least 6 mm
not longer than 20 meters to the ground terminal in the laboratory.
OUTPUT connector
Measurement voltage: 200 V
Max. continuous output current: 50 mA
Max. output frequency: 5 kHz
Red LED
See Table 6-1 "Signal LEDs"
on page 16
SPECTANO 100 front view
peak
STOP button
peak
Stops measurements
Green LED
See Table 6-1 "Signal LEDs"
on page 16
Device description
2
cross-section and
INPUT connector
Yellow LED
See Table 6-1 "Signal LEDs"
on page 16