Typical Setup For A Test Cell Without Guard Ring - Omicron Lab SPECTANO 100 Quick Start Manual

Hide thumbs Also See for SPECTANO 100:
Table of Contents

Advertisement

SPECTANO 100 Quick Start Guide
7.4.2

Typical setup for a test cell without guard ring

26
To connect SPECTANO 100 to a test cell with guard ring:
1. Ground SPECTANO 100 according to the safety rules (see 2.2 "Five safety
rules" on page 5).
2. Connect the standard triaxial cable according to the color code to the
SPECTANO 100 OUTPUT and INPUT connectors:
• Red-labeled cable to the OUTPUT connector
• Blue-labeled cable to the INPUT connector
3. Depending on the test cell connection (BNC or 4 mm plug), connect one of
the adapter cables described earlier in this chapter with the corresponding
color code to the standard triaxial cable:
• Red-labeled adapter cable to the red marked standard cable
• Blue-labeled adapter cable to the blue marked standard cable
4. Connect the adapter cables to the test cell with guard ring according to the
schematic diagram.
Note: If you are using the triaxial-to-4 mm plug adapter cable, be aware that the
inner shield (guard) and the outer shield (ground) are not connected.
For more information and example connections to different types of test cells,
see "Appendix" in the SPECTANO 100 User Manual.
DANGER
Death or severe injury caused by high voltage or current
► Do not touch the OUTPUT connector when the red LED is on or blinks.
► Always ensure that no dangerous voltages are present at any touchable
part of the device, test cell or sample.
► Do not touch the output if the guard (inner shield of the triaxial connection)
is identical with the ground (outer shield of the triaxial connection).