GETTING STARTED
CHAPTER 1
Figure 1-12: Scale Branch Menu Structure
1.9
Calibration
The Calibration screen enables entry of a GEO code adjustment value, linearity adjustment, and
provides access to zero and span adjustments.
1.9.1 GEO Code
Enter the GEO code for the appropriate adjustment value for the current geographical location.
GEO codes are numbered 0–31. (Refer to Appendix E.)
Linearity Adjustment
Select if a linearity adjustment will be used during calibration. Selections are as follows:
Disabled – Use only zero and one span point
Enabled – Use zero, mid span point, and high span point
Ohaus Corporation www.ohaus.com
1-18
7000 Series T72XW Indicator Service Manual