STARTER 5000 Bench pH Meter EN-11 STARTER 3100C Bench Conductivity Meter Instruction Manual Manual de instrucciones del Conductivímetro de mesa STARTER 3100C STARTER 3100C Appareil de mesure de la conductivité de laboratoire Manuel d’instructions STARTER 3100C Medidor de Condutividade de Bancada Manual de Instruções...
Set temperature unit ................8 Set MTC temperature value .............. 8 Set calibration standard ..............8 Set temperature correction coefficient ..........8 Set TDS factor ................... 9 STARTER 3100C OPERATION .......... 10 Calibration ..................10 4.1.1 Selecting a standard ..............10 4.1.2 Performing a calibration ............
Thank you for choosing OHAUS. Please read the manual completely before using the STARTER 3100C bench conductivity meter to avoid incorrect operation. The STARTER 3100C has an excellent performance/price ratio and is designed with many useful features. Starter 3100C offers many practical features such as:...
EN-2 STARTER 3100C Bench Conductivity Meter Safety Precautions CAUTION: Read all safety warnings before installing, making connections, or servicing this equipment. Failure to comply with these warnings could result in personal injury and/or property damage. Retain all instructions for future reference.
STARTER 3100C Bench Conductivity Meter EN-3 Display and controls Displays Endpoint stability / Auto endpoint Measurement icon - measurement or calibration is running Calibration icon - calibration in progress Setup icon - instrument is in the setup mode TDS factor in the setup mode...
Page 6
EN-4 STARTER 3100C Bench Conductivity Meter Controls Button Press & release Press & hold for 3 seconds - Start or endpoint measurement - Turn auto endpoint on / off - Confirm setting, store entered value - Start calibration - Review the latest calibration...
STARTER 3100C Bench Conductivity Meter EN-5 INSTALLATION Carefully unpack the meter. Package contents The ST3100C-B package has the following items ST3100C-B Units STARTER 3100C meter Stand alone electrode holder 1413 μS/cm standard solution 1 bottle (about 20ml) 12.88 mS/cm standard solution...
Installing the power adapter Insert the right adapter clip into the power adapter slot. The Ohaus balance power supply is also 12V which could be used for the pH meter. GlobalTestSupply www.
There is one socket “COND.” for conductivity electrode. Meter stand for adjusting view angle One unique design of the STARTER 3100C is the meter stand, which can be used to adjust the view angle of the display in case of working on high lab table.
EN-8 STARTER 3100C Bench Conductivity Meter SETUP Set temperature unit Power the meter on by pressing Press and hold until the set up icon appears on the display and the current temperature unit blinks. (°C or °F) buttons to select temperature unit .
STARTER 3100C Bench Conductivity Meter EN-9 buttons to adjust the temperature correction value. (0.00 to 10.00 %/°C). Press to confirm your setting. Continue with reference temperature setting by following or press to go back to the measurement screen. Note: If you set the temperature correction coefficient value as 0 %/°C , that means NO Temperature Compensation for the conductivity measurement.
Calibration 4.1.1 Selecting a standard When using the STARTER 3100C conductivity meter, you have to select a standard for calibration (see 3.3). Press and hold the enter setup mode, after press to confirm the temperature unit and value.
STARTER 3100C Bench Conductivity Meter EN-11 When in Auto End Point Mode, the meter determines when the reading is stable then displays and locks the reading or calibration value automatically, the reading freezes and blinks 3 times then disappears; blinks 3 times and freezes on the display.
Printing If a printer (e.g. SF-F40A) is connected to the STARTER 3100C, a print-out is automatically generated after each end pointed measurement or calibration. (printer baud rate 9600bps; 8 data bit;...
Page 15
STARTER 3100C Bench Conductivity Meter EN-13 The details for the second line are: Auto EP, 85.1μS/cm, 26.8°C, ATC |---- Auto Temp. Compensation |----------------Temperature value and unit |------------------------- Conductivity value |---------------------------------------- Auto End Point The details for the third line are: Ref.25°C , T.Comp.Coeff.
Meter maintenance Never unscrew the two halves of the housing! The STARTER 3100C does not require any maintenance. To clean, use a damp cloth. The housing is made of acrylonitrile butadiene styrene (ABS). This material is susceptible to damage by some organic solvents, such as toluene, xylene and methyl ethyl ketone (MEK). Any spillage should be immediately wiped off.
STARTER 3100C Bench Conductivity Meter EN-15 Note: You have to finish pressing all five keys within two minutes, otherwise Err 1 appears and you will have to repeat the procedure. Recover Factory Setup When the meter is off , press and hold &...
EN-16 STARTER 3100C Bench Conductivity Meter TECHNICAL DATA Specifications Ambient conditions Indoor use only Altitude: Up to 2000 m Specified Temperature range: 5°C to 40°C Humidity: maximum relative humidity 80 % for temperatures up to 30°C decreasing linearly to 50% relative humidity at 40°C Mains supply voltage fluctuations: up to ±10% of the nominal voltage...
ISO 9001 Registration In 1994, OHAUS Corporation, USA, was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International (BVQI), confirming that the OHAUS quality management system is compliant with the ISO 9001 standards requirements. On June 21, 2012, OHAUS Corporation, USA, was re-registered to the ISO 9001:2008 standard.
Page 21
Ohaus. In lieu of a properly returned warranty registration card, the warranty period shall begin on the date of shipment to the authorized dealer.