Calibration Menu; Initial Calibration; Zero Calibration [Zero]; Span Calibration [Span] - OHAUS Dafender 3000 i-DT33P Instruction Manual

Table of Contents

Advertisement

EN-28
4.2

Calibration Menu

Enter the calibration menu

Initial Calibration

4.2.1
When the scale is operated for the first time, a zero and span calibration are recommended to ensure accurate
weighing results.
Before performing the calibration, be sure to have the appropriate calibration weights as listed in table 4-1.
Ensure that the LFT switch/calibration lock is set to the unlocked position.
You can also adjust the GEO setting according to your location
the Geo Code Adjustment section below.
Max Capacity (kg)
Note: The calibration unit is consistent with the capacity unit.
4.2.2

Zero Calibration [ZErO]

Zero calibration uses one calibration point. The zero calibration point is established with no weight on the scale.
Use this calibration method to adjust for a different static load without affecting the span or linearity calibration.
Calibration procedures:
1.
Long press the Menu button until you see
2.
Release the button and wait for the display to show
3.
The display shows
0
4.
kg and the calibration unit are flashing on the display. With no weight on the pan, press the Yes button to
establish the zero point.
5.
The display shows
Note:
If zero calibration is failed or if after 40 seconds the calibration is still not successful,
3 seconds and the previous calibration data will be restored.
6.
Then the display shows

Span Calibration [SpaN]

4.2.3
Span calibration uses one calibration point. It is established with a calibration mass placed on the scale.
Note: Span calibration should be performed after zero calibration.
Calibration procedures:
1.
Long press the Menu button until you see
2.
Release the button and wait for the display to show
3.
Short press the No button to navigate until you see
4.
The calibration point and the unit are flashing on the display based on the scale capacity and unit set in the
030.000
menu. (e.g.
5.
To change the calibration point:
a)
Short press the No button several times until the desired number appears.
Note: press the Back button can decrease the digit.
b)
Short press the Yes button to accept the number and move on to the next digit.
C.A.L
to perform calibrations.
Table 4-1 Suggested Span Calibration Mass (Sold Separately)
Mass
3 kg
3 kg
6 kg
6 kg
15 kg
15 kg
30 kg
30 kg
60 kg
60 kg
150 kg
150 kg
ZErO.
Press the Yes button.
--C--
-DONE-
, and then
SPAN
. Press the Exit button to exit.
kg) If you do not need to change the calibration point, skip to step 6.
.
For detailed GEO information, please check
*
(kg)
Max Capacity (lb)
6 lb
12 lb
30 lb
60 lb
120 lb
300 lb
mM.E.N.U
.
C.A.L
. Press the Yes button.
when the Zero calibration is finished.
mM.E.N.U
.
C.A.L
. Press the Yes button.
SPAN.
Press the Yes button.
Defender 3000 Indicators
Mass* (lb)
6 lb
12 lb
30 lb
60 lb
120 lb
300 lb
CAL E
is displayed for

Advertisement

Table of Contents
loading

Table of Contents