Marquette Hellige GmbH
Electrical Diagram for Patient Leakage Current Test
For the protection of the test person, the following values of resistor R may be used:
Type BF
Type CF
7.2
Maintenance, cleaning, disinfection
MAC 1200 maintenance, cleaning, disinfection is performed in accordance with the
MAC 1200 "Operator's Manual", Section 11, "Cleaning, disinfection and maintenance",
as applicable.
Applications requiring extensive recordings may result in deposits on the thermal array
printhead which normally do not, however, have any adverse effect on the printing
quality. This can be removed with a soft, fluff-free cloth soaked in an alcohol-based
cleaning agent (e.g., surgical spirit).
MAC 1100/1200 V 1.1
227 492 20 D - 0002
22 kOhm (120 to 130 V)
47 kOhm (220 to 240 V)
100 kOhm (220 to 240 V)
Page 58