(15)
Printing Function
1) Printing Method
Graphic print by a built-in thermal printer
2) Printed Data
(i)
Stored sample data (the same as the measured data)
(ii) Quality control (QC data, QC chart)
(iii) Measured data (Date, Time, Sample ID number, Parameter name, Measured data, Reaction
curve, Analysis print data, Rack number (CA-550 and CA-560 only))
(iv) Standard curve (SC data, SC chart)
(v) Confirmation messages, Error messages
(vi) Maintenance information and various setting values (CA-550 and CA-560 can print at each
parameter.)
(16)
Quality Control Function
1) L-J control or X control is possible using control material.
2) Applicable parameters:
3) Number of stored files:
4) Number of data points:
(17)
Select Function
1) Measurement Mode
(i)
Random access measurement mode
5 parameters are selected out of 7 parameters (14 parameters in CA-550 and CA-560) and the
five parameters are measured at random.
(ii)
Programmable measurement mode
All parameters are measured with changing the measurement order.
(iii)
Replication mode
The same sample is measured twice (or more) and the mean value is determined to be the
measurement result. However, the throughput is less than half number of that of ordinary
measurement.
2) Settings
(i)
Sample ID number
(ii)
Date/Time
(iii)
Setting function for APTT heating time
2, 3, 4 or 5 (minutes) of heating time can be set. (Standard setting is 3 minutes and the
throughput changes if other time is set.)
3) Output
Automatic transfer or Manual transfer can be set.
Raw data output is available in Service mode. (CA-550 and CA-560 only)
4) Stored data processing
Displaying and printing the stored data is possible.
5) Service Function
For the Customer:
For the Servicing Personnel:
CA-500 Series S/M
PT, APTT, Fbg, TTO, NT, TT, AT III
14 parameters (CA-550 and CA-560 only)
-Fbg, +Fbg, +DD, +DDPl, +SFD, and +PFD are also treated as
parameters and QC analysis can be performed on each parameter.
(CA-550 and CA-560 only)
12 files for each parameter
67 files (CA-550 and CA-560 only)
180 for each parameter
Displaying of standard curve data and setting of abnormal values
(range)
Service mode, System tests, Memory Initialization, and
adjustment/setting (of X-Y axis position)
1-12
Revised December 2001
8