Catcher Test - SYSMEX CA-500 Service Manual

Automated blood coagulation analyzer
Table of Contents

Advertisement

(2) To verify Tube Transfer operation
1)
Transfer the tube of the specified tube rack in the following order.
incubation well
2)
DO NOT leave reaction tubes on the Reaction Tube Rack to prevent malfunction.
[ ], [ ] keys: Use to move cursor.
[+] key:
[-] key:
[Set] key:
[Return] key: Press to quit the test and to return to Test Menu display.
(3) To cancel tube transfer test operation
Press [Cancel] key to stop the operation, then the tube transfer test setting screen will appear.
5.5.6

Catcher Test

Verify the catcher operation.
(1) Press [Special Operate]
screen will be displayed.
(2) To verify catcher operation
Set reaction tubes in the "1", "5", "21", "25", "31", "35", "56" and "60" position of the Reaction Tube Rack.
DO NOT place tubes in the other positions.
[RESET] key: Use to clear the test cycle counts.
[Set] key:
Press to perform the catcher test.
[Return] key: Press to quit the test and to return to Test Menu display.
(3) To cancel catcher test operation
Press [Cancel] key to stop the operation, then the catcher test setting screen will appear.
CA-500 Series S/M
reagent dispense
Use to increase value on which cursor is positioned.
Use to decrease value on which cursor is positioned.
Press to perform the tube transfer test.
[Service]
[Test Menu]
Ready
Sysmex
Replace Rack? YES!
Catcher Test
Tube
No. = 5
Warm
No. = 1
Detect
No. = 1
Test Stop
0
/
Set
Figure 5-5-7: Catcher Test Screen
detector well
discard
[Catcher Test] keys to start, then the following
IP
0
Return
RESET
5-13
Revised December 2001
8