The changeable parameters are as follows:
Parameter
Description
Vial Name
Name of Reagent Vial
Std
Normal Volume (mL)
Dead
Dead Volume (mL)
2
Area [mm
]
Section Area of Vial (mm
Offset
Number of Offset Pulse
from Adjustment Position
to Dead Volume
[Set] Key
Use to execute Descend Pulse Adjustment screen.
[Cancel] Key
Cancels the Z Pulse Down Adjustment and returns to the Reagent Vial Settings screen.
<Z Pulse Down Adjustment>
(1) Fill the dead volume of the reagent in the reagent vial and place at the reagent holder No. 9.
(2) Press [Set] key to execute Z Pulse Down Adjustment.
(3) The probe moves to the reagent holder No. 9 and detect a liquid surface for 5 times.
(4) The difference between the average of the descend pulse and the value of the reagent holder No. 9
position adjustment will be calculated as the pulse correction value.
CA-500 Series S/M
Key to be Used
Full Keyboard
Numeric Keys
Numeric Keys
2
) Numeric Keys
Numeric Keys
Ready
Sysmex
Replace Rack? YES!
Reag. Vial Settings
Vial Name
Cup
4.00 mL
Std
0.10 mL
Dead
Area [mm ]
105.683
2
Offset
0
Adjust Z pulse down
Set
Figure 5-7-3: Z Pulse Down Adjustment Screen
5-20
Changeable Range
9 characters maximum
(Alphabets and Numbers)
0.01 thru 99.99 (by 0.01 mL)
0.00 thru 99.99 (by 0.01 mL)
0.000 thru 999.999 (by 0.001 mm
-999 thru 999 (by 1 pulse)
IP
Cancel
Revised December 2001
2
)
8