2 Safety and Precautions

2.1 General Safety Tips

Installation of the HATCH X-RAY must be done only by an authorized service engineer. Consult the
factory or dealer as necessary.
Make sure that the HATCH X-RAY is assembled and installed in compliance with all applicable laws and
recommendations concerning electrical safety.
The unit contains and generates high voltages. Only a trained service personnel should attempt to open
the protective plastic covers or repair the unit.
This X-ray equipment may be dangerous to the patient and the operator unless safe exposure factors and
operating instructions are observed. Follow proper X-ray radiation safety rules.
Follow instructions specified in this manual when carrying out exposures during installation.
Do not use non prescribed exposures.
Make sure that only the operator and the patient are present in the treatment room during radiation
exposure procedure.
Always be at a distance of more than 2 meters away from the Tube head while carrying out exposures.
Exercise caution when operating and installing the mechanical suspension arm. The arm is spring
loaded and can bounce out if proper installation procedures are not followed.
Where complete safeguarding of the equipment is not possible, due care must be taken to ensure that no
part of your body or clothing can be trapped or injured by any part of the equipment. In particular, make
sure that fingers are not caught or pinched during scissor arm movement.
Ensure proper electrical grounding. A bad grounding can be dangerous for the operator and can generate
malfunctioning of the device.
Turn Off and Remove All Power Before Performing Any Service. Wait for at least 2 minutes after
mains power off before opening and accessing the covers. During this time, remove the mains plug
from the wall socket or turn off circuit breaker.
Scissor arm can open out during installation of the unit which may cause injury
to persons / patient standing close to the equipment Always make sure to lock
the movement of the scissor arm in folded condition
2.2

Safety Symbols

The following safety related symbols are found on the equipment.
Caution Symbol
This symbol indicates the user to be cautious and refer to the user manual
for safe operating instructions.
Protective Earth Ground
Mains Earth Ground is required for continued protection against shock
hazards.
Type of Insulation
Class 1, Type B Insulation. Protection against electric shock
(UL60601-1:2003). Requires protective Earth Connection.