5.9
Remote Console Installation (Optional)
•
Take the remote console template from packing box and tape it at the desired location on the wall stud.
•
Use the Spirit level and make sure that the template is level.
•
Mark the locations for the 2 mounting holes on the template.
•
Drill the holes marked using 9/64-inch drill bit to the depth of 0.84" as shown in Figure 52.
•
Take the remote console assembly and pull out the console from the slot.
•
Remove the plastic cover by removing 4 ( M3x6 Hex socket head cap) screws with M3 plain washers using a 2.5
mm Allen key as shown in Figure 53 .
•
As shown in Figure 54 , lift the cover along with console assembly and disconnect the console cable from the
board. Keep the parts aside for reuse.
Figure 52
•
Take the remote console and route the remote console communication cable through the hole provided on the
plate as shown in Figure 55.
•
Secure the remote console plate on the stud by inserting the 2 no's of 6.3 x 32mm Self Tap Wood Screw at the
locations shown in Figure 55. Tighten using an 8 mm box spanner.
•
Connect the remote console communication cable as per the connections shown below in 5.9 HATCH X-RAY
Console Configurations.
•
Follow the same steps for the second console as above if double doorbell switch is used.
•
Connect the external console cable to J8 connector and close the cover of the remote console by installing 4 (M3x6
Hex socket head cap) screws using a 2.5 mm Allen key as shown in Figure 53.
•
Secure the remote keypad console in the slot of the cover.
Figure 53
Figure 55
Page 20 of 34
Figure 54