Measure tp:
See point „4.4.2.2.1"
Set:
In the last step, the calibration settings can be saved with "Save" to the previously selected calibration
memory location, or can be canceled with "Discard".
NOTE:
After performing "Save", the original material calibration is preceded by an "OWN:", indicating
that this is a specially prepared material calibration.

4.4.3 Offset Balancing

In order to compensate measurement errors e.g. due to density deviations in the material or due to the
installation conditions, there is an option to perform a linear displacement of the measurement value.
This is the purpose of this menu item. It is possible to displace the measurement value between -10 and
+10 percent points. The set displacement is stored in the probe and will subsequently also affect the
analogue output. The setting is maintained in a traceable manner.
18
Measure tp
The tp value for the upper moisture value can be
adjusted manually using the
button
to move to the previous item.
Use the
"Discard". The corresponding sub-point is selected
with the key
to the previous point.
Adjust the offset to the desired value using the
buttons
the probe with the button
item with the button
/
to accept the setting and press the button
/
buttons to toggle between "Save" and
and the key
/
. Subsequently store the set value in
. You can exit this menu
.
buttons. Press the
can be used to switch