Physical Specifications

Physical Specifications
Item
Environment
Power requirements,
main unit
Power requirements,
power adapter
Dimensions
Weight
Power disconnect
clearance (rear)
Ambient operating
temperature
Ambient storage
temperature
Humidity restrictions
Altitude restrictions
Sound pressure level
(maximum)
Installation category
Pollution degree
Data connection

Electromagnetic Compatibility

Regulatory for Canada (ICES/NMB-001:2006)
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est confomre à la norme NMB-001 du Canada.
ISM Equipment Classification (Group 1, Class A)
This equipment is designated as scientific equipment for laboratory use that intentionally
generate and/or use conductively coupled radio-frequency energy for internal functioning,
and are suitable for use in all establishments, other than domestic and those directly
connected to a low voltage power supply network which supply buildings used for domestic
purposes.
23
Description
Indoor use only
12 VDC, 4 amps (50 VA)
100 VAC to 240 VAC ±10%, 1.5 amps, frequency range: 50/60 Hz
31.50 cm W x 18.20 cm H x 43.50 cm D
(12.40 in. W x 7.17 in. H x 17.13 in. D)
6.60 kg (14.55 lbs)
20 cm to 30 cm (7.9 in. to 11.8 in.) between the rear of the instrument
and the wall
15°C to 40°C (59°F to 104°F)
–20°C to 50°C (–4°F to 122°F)
15% to 85%, non-condensing
Up to 2000 m (6562 ft)
45 dB
II
2
One (1) USB 2.0 port