Emerson Rosemount 396P Quick Start Manual page 35

Ph/orp sensors
Hide thumbs Also See for Rosemount 396P:
Table of Contents

Advertisement

November 2020
Figure 2-17: Wiring for Rosemount 396PVP and Rosemount 5081-P-HT
If there is a shrink sleeve on the gray reference wire, remove it before
connecting the wire to the terminal.
Table 2-17: Wiring for Rosemount 396PVP and Rosemount 5081-P-HT
Terminal
number
N/A
1
2
3
4
5
6
7
8
Quick Start Guide
Wire color
Connects
to
Green
N/A
N/A
N/A
Clear
N/A
White
RTD return
White/red
RTD sense
Red
RTD in
White/gray
Drain
Gray
Reference
Blue
Solution
ground
Terminal
Wire color
number
9
Clear
10
Orange
11
N/A
12
N/A
13
N/A
14
N/A
15
N/A
16
N/A
Quick Start Guide
Connects
to
Drain
mV in
-5 Vdc
+5 Vdc
Anode
Cathode
-24 Vdc
+24 Vdc
35

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rosemount 396pvpRosemount 396pvp-70

Table of Contents