Quick Start Guide
Figure 2-2
-05.
Figure 2-2: Wiring for Rosemount 396P (Blue Cable) and Rosemount
1056, 56, or 1057
If there is a shrink sleeve on the gray reference wire, remove it before
connecting the wire to the terminal. For Rosemount 1056, terminate inner
drain with a wire nut.
Table 2-2: Wiring for Rosemount 396P (Blue Cable) and Rosemount 1056, 56, or
1057
Terminal number
1
2
3
4
5
6
7
8
9
10
N/A
20
and
Table 2-2
are applicable to option codes -01, -03, -04, and
Wire color
White
White/red
Red
Blue
Inner drain
White/gray
Shield
Orange
N/A
Gray
Green
November 2020
Connects to
RTD return
RTD sense
RTD in
Solution ground
+5 Vdc
-5 Vdc
pH shield
pH in
Reference shield
Reference
N/A