November 2020
Figure 2-7
and -06.
Figure 2-7: Wiring for Rosemount 396P (Gray Cable) and Rosemount
1066
Note
•
If ground lead is present, terminate it to green ground screw on inner
enclosure.
•
TB5, TB7, and TB8 not used for pH/ORP sensor wiring.
Table 2-7: Wiring for Rosemount 396P (Gray Cable) and Rosemount 1066
Terminal number
TB3
TB3
TB3
TB4
TB4
TB1
TB1
TB2
TB2
TB2
Quick Start Guide
and
Table 2-7
are applicable to option codes -01, -03, -04, -05,
Wire color
White
White/red
Red
Green
Brown
Black/white
Black
Blue
White/gray
Gray
Quick Start Guide
Connects to
RTD return
RTD sense
RTD in
Preamplifier: +volts
Preamplifier: -volts
pH shield
pH in
Solution ground
Reference shield
Reference in
25