Emerson Rosemount 396P Quick Start Manual page 29

Ph/orp sensors
Hide thumbs Also See for Rosemount 396P:
Table of Contents

Advertisement

November 2020
Figure 2-11: Wiring for Rosemount 396PVP and Rosemount 1066
Note
If ground lead is present, terminate it to green ground screw on inner
enclosure.
TB5, TB7, and TB8 not used for pH/ORP sensor wiring.
Table 2-11: Wiring for Rosemount 396PVP and Rosemount 1066
Terminal block
TB3
TB3
TB3
TB1
TB1
TB2
TB2
TB2
Quick Start Guide
Wire color
White
White/red
Red
Braid or clear
Orange
Blue
White/gray
Gray
Quick Start Guide
Connects to
RTD return
RTD sense
RTD return in
pH input - pH shield
pH input - pH in
Solution ground
Reference shield
Reference in