Emerson Rosemount 396P Quick Start Manual page 22

Ph/orp sensors
Hide thumbs Also See for Rosemount 396P:
Table of Contents

Advertisement

Quick Start Guide
Figure 2-4
Figure 2-4: Wiring for Rosemount 396P (Blue Cable) and Rosemount
1056, 56, or 1057
If there is a shrink sleeve on the gray reference wire, remove it before
connecting the wire to the terminal. For Rosemount 1056, terminate inner
drain with a wire nut.
Table 2-4: Wiring for Rosemount 396P (Blue Cable) and Rosemount 1056, 56, or
1057
Terminal number
N/A
1
2
3
4
5
6
7
8
9
10
N/A
22
and
Table 2-4
are applicable to option codes -02, -07, and -08.
Wire color
Shield
White
White/red
Red
Blue
N/A
N/A
Braid
Orange
White/gray
Gray
Green
November 2020
Connects to
N/A
RTD return
RTD sense
RTD in
Solution ground
+5 Vdc
-5 Vdc
pH shield
pH in
Reference shield
Reference
Ground
Emerson.com/Rosemount