November 2020
Figure 2-9
Figure 2-9: Wiring for Rosemount 396P (Gray Cable) and Rosemount
1066
Note
If ground lead is present, terminate it to green ground screw on inner
enclosure.
TB5, TB7, and TB8 not used for pH/ORP sensor wiring.
Table 2-9: Wiring for Rosemount 396P (Gray Cable) and Rosemount 1066
Terminal number
TB3
TB3
TB3
TB1
TB1
TB2
TB2
TB2
Quick Start Guide
and
Table 2-9
are applicable to option codes -02, -07, and -08.
Wire color
White
White/red
Red
Braid or clear
Orange
Blue
Braid or clear
Gray
Quick Start Guide
Connects to
RTD return
RTD sense
RTD return in
pH shield
pH in
Solution ground
Reference shield
Reference in