Figure 2-12: Wiring for Rosemount 396PVP-70 and Rosemount 1066
Note
If ground lead is present, terminate it to green ground screw on inner
enclosure.
TB5, TB7, and TB8 not used for pH/ORP sensor wiring.
Table 2-12: Wiring for Rosemount 396PVP-70 and Rosemount 1066
Terminal block
TB3
TB3
TB3
TB4
TB4
TB1
TB1
TB2
TB2
30
Wire color
White
White/red
Red
Inner drain
White/gray
Shield
Orange
Blue
Gray
November 2020
Connects to
RTD return
RTD sense
RTD return in
Preamplifier: +volts
Preamplifier: -volts
pH shield
pH in
Solution ground
Reference in