Ph/orp sensors
Hide thumbs Also See for Rosemount 396P:
Table of Contents

Advertisement

November 2020
Figure 2-5: Wiring for Rosemount 396PVP without Integral Preamplifier
and Rosemount 1056, 56, or 1057
If there is a shrink sleeve on the gray reference wire, remove it before
connecting the wire to the terminal. For Rosemount 1056, terminate inner
drain with a wire nut.
Table 2-5: Wiring for Rosemount 396PVP without Integral Preamplifier and
Rosemount 1056, 56, or 1057
Terminal number
N/A
1
2
3
4
5
6
7
8
9
10
Quick Start Guide
Wire color
Clear
White
White/red
Red
Blue
N/A
N/A
Clear
Orange
Clear
Gray
Quick Start Guide
Connects to
N/A
RTD return
RTD sense
RTD in
Solution ground
+5 Vdc
-5 Vdc
pH shield
pH in
Reference shield
Reference